CM-3610A

CM-36dG

World’s first Colour-and-Gloss Bench-top spectrophotometer with versatile horizontal or top-port alignment. Innovative measurement technology and best price/performance ratio in the field of reflectance and transmission colour measurements.

  • Bench-Top Instrument
  • Wavelength range 360-740nm
  • d:8° Sphere Geometry
  • Gloss Measurement 60°
  • Reflection & Transmission
  • 4 Apertures SAV / MAV / MLAV / LAV
CM-3610A

High-precision spectrophotometer with ISO compliant gloss sensor and stability check

The CM-36dG Spectrophotometer is a high-precision and high-reliability bench-top instrument capable of measuring colour either in reflectance or transmittance, ideally suited to a wide range of applications such as Plastics, Paints, Ceramics, Chemicals etc. 
For the first time ever, Konica Minolta Sensing has integrated an ISO 2813 compliant 60° gloss sensor inside a benchtop spectrophotometer, to simultaneously measure Colour and Gloss. As both values are measured and reported together operator errors are reduced, the Quality Control workflow can be streamlined and equipment and maintenance costs saved.
The CM-36dG utilizes the patented and proven Numerical UV Control (NUVC), unsurpassed technology for UV adjustments when measuring samples that contain optical brighteners such as pulp, paper, textiles or chemicals.
Several new features have been included to improve the user experience, status LEDs provide clear visual feedback, a camera preview system for sample positioning and reporting, and versatile port alignment allowing the device to be used horizontally or rotated 90° to measure e.g. powdery materials in “Top-port” style.
Konica Minolta’s patented technologies assure the highest accuracy and repeatability levels for this class of instrument. Konica Minolta’s renowned accuracy and stability is supported by our latest innovation “Wavelength Analysis and Adjustment” (WAA) to give customers the ability to check and adjust the CM-36dG during calibration, assuring that the instrument is in perfect condition.  

 

Specifications & Downloads

Specifications & Downloads
Model Spectrophotometer CM-3610A
Illumination / Viewing system Reflectance: d:8° (diffused illumination, 8-degree viewing) Conforms to ISO 2469, JIS P8148, DIN 53145-1 and DIN 53145-2 standards.
Light receiving element Silicon photodiode array (dual 40 elements)
Spectral separation device Diffraction grating
Wavelength range, pitch 360 nm to 740 nm
Wavelength pitch 10 nm
Reflectance range 0-200%, resolution 0.01%
Light source Pulsed xenon lamps (× 3)
Measurement time Approx. 1.5 seconds (for measurements of fluorescent colors, at 9600 bps)
Minimum interval between measurements Approx. 4 seconds; when reflectance measured, approx. 5 seconds; when fluorescent color
Measurement/illumination area Ø 30 mm / Ø 34 mm
Inter instrument agreement Mean ΔE*ab 0.2 based on 12 BCRA Series II color tiles compared to values measured with master body
Repeatability Spectral reflectance: Standard deviation within 0.1% Colorimetric values: Standard deviation within ΔE*ab 0.02 (condition; white calibration plate measured 30 times at 10-second intervals)
Temperature drift Spectral reflectance: Within +/- 0.10%/°C Color difference: Within ΔE*ab 0.05/°C
UV adjustment Instantaneous numerical adjustment
UV cut filter 400 nm cutoff and 420 nm cutoff
Control method Directly connected to a computer
Interface RS-232C format
Power 100-240V AC 50-60Hz 25W AC (with a dedicated AC adapter)
Dimensions (W x H x D) 300 × 585 × 315 mm
Weight 15.5kg
Operating temperature/humidity range 13 - 33°C, relative humidity 80% or less (at 30°C) with no condensation
Standard accessories White calibration plate CM-A 133, Zero calibration box CM-A119, AC adapter AC-A12, RS-232C cable IF-A12, Accessory Case CM-A117, Dust Cover CM-A118, Unit Driver CM-A108